2680A/2686A
Users Manual
4-34
Fluke DAQ Totalizer Count and Direction Test
1.
Configure DIO
In the Configuration dialog, select the DIO from TreeView.
In the DIO Configuration dialog, set the start count to 0, deselect debounce (if
selected), and select direction as down.
2.
Start Spy
Go to the Spy and start the Spy with the play button.
3.
Verify Totalizer Value
Verify that the totalizer value is has not changed.
4.
Set Signal Generator
Set the signal generator to 10 Hz and 1 V pp and
connect to the totalizer input.
5.
Verify Totalizer Count
Verify that the totalizer count decreases by about 10
counts per second.
6.
Stop Spy
Stop the Spy.
7.
Disconnect Signal Generator
Disconnect the signal generator.
Fluke DAQ Totalizer Count and Debounce Test
1.
Configure DIO
In the Configuration dialog, select the DIO from TreeView.
In the DIO Configuration dialog, set the start count to 0, select debounce on,
and select direction as up.
2.
Start Spy
Go to the Spy and start the Spy with the play button.
3.
Verify Totalizer Value
Verify that the totalizer value has not changed.
4.
Set Signal Generator
Set the signal generator to 1 kHz and 1 V pp and
connect to the totalizer input.
5.
Verify Totalizer Count
Verify that the totalizer count does not change by
more than 1 count.
6.
Stop Spy
Stop the Spy.
7.
Disconnect Signal Generator
Disconnect the signal generator.
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...