Preparing for Operation
Instrument Preparation
2
2-35
Reviewing and Setting the Network Type
Perform the procedure below to review or set the network type to isolated. Perform
the procedure in Table 2-11 to review or set the network type to general. An
isolated network consists of only 2680 Series instruments and one or more host
computers. A general network consists of instruments, host computers, and
possibly servers, routers, gateways, or other network devices. (See Figure 2-16 for
examples.)
If you use Fluke DAQ software for isolated network operation and set the
instruments’ network type to isolated, you do not need to know or set IP addresses
for your instruments.
Table 2-10. Reviewing and Setting the Network Type
F
Press the COMM key to review the network type, or press and hold the COMM
key for 3 seconds to set the network type.
C
B
Press the up/down arrow keys until nEt (Network) appears in the primary
display (COMM appears in the secondary display).
I
Press the ENTER key. nEt appears in the secondary display and ISo (isolated
network) or gEn (general network) is in the primary display.
C
B
To set the network type to isolated, press the up/down arrow keys to select ISo
(current setting appears bright).
I
Press the ENTER key to exit. (Pressing any other function key will cancel set
operations.)
Summary of Contents for 2680A
Page 8: ......
Page 14: ...2680A 2686A Users Manual vi ...
Page 20: ...2680A 2686A Users Manual 1 2 ...
Page 98: ...2680A 2686A Users Manual 2 62 ...
Page 196: ......
Page 228: ...2680A 2686A Users Manual A 32 ...
Page 236: ...2680A 2686A Users Manual C 4 ...
Page 242: ...2680A 2686A Users Manual D 6 ...
Page 274: ...2680A 2686A Users Manual 6 ...