Test Specifications
The accuracy specification for the display range is
defined as
±
(%r digit counts) at 23
°
C
±
5
°
C,
≤
75 % RH. Between 0
°
C and 18
°
C and between 28
°
C
and 40
°
C, accuracy specifications may degrade by 0.1 x
(accuracy specification) per
°
C.
The measurement range meets the service operating
errors specified in EN61557-1: 1997, EN61557-2: 1997 ,
EN61557-4: 1997, DIN VDE0404-2.
Power-on Test
The test indicates reversed L-N, missing PE, and
measures the mains voltage and frequency.
Operational Error Measurement Range .......195 V to 253 V
Display Range ...............................................90 V to 264 V
Accuracy at 50 Hz..................................
±
( 2% + 3 counts)
Resolution...................................................................0.1 V
Input Impedance ....................................... > 1 M
Ω
// 2.2 nF
Maximum Input Mains Voltage...................................264 V
Earth Bond Test
Operational Error Measurement Range ..........0.2 to 1.99
Ω
Operational Error ...................................................... 10.0%
Accuracy (after Bond Test zeroing)........
±
( 5% + 4 counts)
Display Range .................................................0…19.99
Ω
Resolution................................................................ 0.01
Ω
Test Current................... 200 mA ac -0% +40% into 1.99
Ω
25 A ac
±
20 % into 25 m
Ω
at 230 V
Open Circuit Voltage.............................> 4 V ac, < 24 V ac
Bond Test Zeroing ...................... can subtract up to 1.99
Ω
Used Current for Bond Test Zeroing .............................10A
Insulation Test
Operational Error Measurement range .............0.1 to 5 M
Ω
Operational Error ........................................................ 9.0%
Accuracy ...................
±
(5% + 2 counts) from 0.1 to 50 M
Ω
±
(10% + 2 counts) from 50 to 299 M
Ω
Display Range .................................................0 to 299 M
Ω