Fluke 6500
Users Manual
38
Resolution ............................................ 1 VA (0 to 999 VA)
0.1 kVA (1.0 kVA to 3.2 kVA)
The appliance under test is energized at mains potential.
Load/Leakage Test: Leakage Current
Operational Error Measurement Range ... 0.25 to 19.00 mA
Operational error....................................................... 12.0%
Accuracy .................................................
±
(4% + 5 counts)
Display Range ......................................... 0.25 to 19.99 mA
Resolution..............................................................0.01 mA
The appliance under test is energized at mains potential.
PELV Test
Display Range ...........................................10.0 V to 39.9 V
Resolution...................................................................0.1 V
Accuracy at 50 Hz..................................
±
( 2% + 3 counts)
Overload protection ............................................. 300 Vrms
Warning threshold..................................................25 Vrms
IEC Lead Test
Earth bond limit...........................................................1.0
Ω
Test current............................................................10 A AC
Insulation resistance limit.........................................
2
.0 M
Ω
Test voltage .........................................................500 V DC
Fuse test and polarity check
Variation Factor Errors
Variation
Factor
Designation %
Variation
Error
Position E1
0.0%
Supply Voltage
E2
5.0%
Temperature E3 5.5%
Current
Consumption
E4 1.5%
Magnetic Fields
E5
2.5%
Impedance E6 1.0%
Capacitance E7 2.0%
Current
Waveshape
E8 1.0%