Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with fluid, installing the probe and connecting power. Consult the
Installation
section for detailed instructions for proper installation of the bath. Be
sure to place the bath in a safe, clean and level location.
Fill the bath tank with an appropriate liquid. For operation at moderate bath
temperatures, clean distilled water works well. Carefully pour the fluid into the
bath tank through the large rectangular access hole above the tank avoiding
spilling any fluid. The fluid must not exceed a height of 1/2 inch below the bath
lid.
The control probe must be inserted through the lid into the bath and plugged into
the socket at the back of the bath.
Power
Plug the bath power cord into a mains outlet of the proper voltage, frequency,
and current capability (see the
Specifications
section). Refer to and read the
Caution at the front of the manual concerning brownout and over voltage
protection.
W
Caution
Check the back panel label for the correct voltage and
frequency prior to energizing the unit.
Set the “HEATER” switch on the front panel to position “LOW” and turn the bath
on using the front panel “POWER” switch. The bath will turn on and begin to heat
or cool to reach the previously programmed temperature set-point. The front
panel LED display will indicate the actual bath temperature.
Set the Temperature
In the following discussion a solid box around the word SET, UP, EXIT or DOWN
indicates the panel button while the dotted box indicates the display reading.
Explanation of the button or display reading are to the right of each button or
display value.
To view or set the bath temperature set-point proceed as follows. The front panel
LED display normally shows the actual bath temperature.
Bath temperature display
When “SET” is pressed the display shows the set-point memory that is currently
being used and its value. Eight set-point memories are available.
Access set-point selection
Set-point 1, 25.0
°
C currently used
24.68 C
1. 25.0