background image

Fluke 707Ex 
Calibration Information
 

Recommended Test Equipment 

A list of recommended equipment for the 

performance verification tests and calibration 

adjustment procedure is shown in Table 3. 

Performance Verification  

XW

 Warning 

To avoid electric shock, do not perform 
the verification tests unless the mA 
Calibrator is fully assembled. 

Perform the following tests to check the accuracy of 

each of the mA Calibrator’s functions against its 

specifications. 

If the mA Calibrator fails any of these tests, 

calibration adjustment or repair is required. 

Performance Verification Procedures 

DC Current Source Mode 

Test 1:  

1.

 

Turn the mA Calibrator on. 

2.

 

After the P.S. power timeout displays, the mA 

Calibrator should display 4.000 mA. If it is 

displaying 0.000 mA, turn the mA Calibrator off. 

Hold down 

A

 while turning the mA Calibrator 

on. Continue to hold 

A

 for 2 seconds. When the 

timeout display ends, the mA Calibrator should 

display 4.000 mA. 

3.

 

Connect the mA Calibrator’s 

+

 terminal to the HP 

3458A 

I

 input, and the 

COM

 terminal to the HP 

3458A 

LO

 input.  

4.

 

Verify that the mA Calibrator's display shows 

SOURCE

 in the upper-left corner.  

5.

 

Set the HP 3458A to measure DC Amps (DCI). 

6.

 

Refer to Table 4 to verify the readings on the HP 

3458A for the following tests. (No adjustment is 

necessary for test 1.) 

Test 2:  

1.

 

Press 

B

 

twice to select 12.000 mA.  

2.

 

Verify the readings. 

Test 3:  

1.

 

Press 

B

 

twice to select 20.000 mA.  

2.

 

Press and rotate the knob until 24.000 mA is 

displayed.  

3.

 

Verify the readings. 

DC Current Measurement Mode 

Tests 4-6: 

1.

 

Press the 

A

 key twice to put the mA Calibrator 

into the mA measurement mode.  

2.

 

The mA Calibrator’s display should show 

MEASURE 

in the top right corner. 

3.

 

Disconnect the HP 3458A from the mA 

Calibrator. 

4.

 

Connect the test leads from the 

red AUX

 terminal 

of the Fluke 5520A to the HP 3458A 

I

 input.  

5.

 

Connect a test lead from the 

black AUX

 terminal 

of the Fluke 5520A to the 

COM 

terminal on the 

mA Calibrator.  

6.

 

Connect a test lead from the HP 3458A 

LO

 input 

to the mA Calibrator's 

+

 

terminal. 

7.

 

Adjust the Fluke 5520A to output the value of test 

4 in Table 4 as measured by the HP 3458A. 

8.

 

Verify the display readings on the mA Calibrator. 

9.

 

Repeat adjustment and reading for steps 5 and 6. 

Table 3. Recommended Equipment 

Equipment 

 

Recommended Model or Equivalent 

DC Calibrator 

DC Voltage: 0 to 30 V 
Accuracy: 

±

 0.004 % +0.5 mV 

DC Current:  0 to 24 mA  
Accuracy: 

±

 0.004% +0.5 

μ

Fluke 5520A Multi-Product Calibrator 

Digital Multimeter 

DC Current: 0 to 26 mA  
Accuracy: 

±

 0.004% +0.5  

μ

HP 3458A 

Flexible Test Leads  

 

Fluke TL24 

Summary of Contents for 707Ex

Page 1: ...ete operating instructions refer to the 707Ex Users Manual located on the CD that came with the mA Calibrator Warranty This Fluke product comes with a 3 year warranty for material and workmanship For warranty information refer to the 707Ex Users Manual that came with the mA Calibrator Contacting Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE...

Page 2: ... to 18 C and 28 C to 50 C Relative Humidity 95 up to 30 C 75 up to 40 C 45 up to 50 C Vibration Random 2 g 5 to 500 Hz Shock 1 meter drop test Safety Compliance Complies with IEC 61010 1 95 CAT I 28 V CSA C22 2 No 1010 92 NRTL ANSI ISA S82 02 01 1994 Directive 94 9 EG and p NEC 500 Uo 27 6 V Io 96 13 mA Co 76 nF Lo 2 5 mH Ui 30 V Ii 24 mA Ci 10 nF Li 0 mH CE Complies with EN61010 1 and EN61326 Pow...

Page 3: ...t is used Measurement transients are defined in IEC1010 1 as 2 µs rise time with a 50 µs duration at 50 of the maximum amplitude height Measurement Category I CAT I is defined for measurements performed on circuits not directly connected to the mains Do not use in a damp or wet environment Faults and Damage XW Warning Applying a voltage greater than 28 V to the input of the mA Calibrator invalidat...

Page 4: ...0460 2 P Conforms to European Union requirements F Direct current Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Conforms to relevant Australian standards Maintenance XW Warning To avoid electric shock personal injury or damage to the mA Calibrator Do not service this product To maintain the integrity of the mA Calibrator in explosive atm...

Page 5: ...or handling them 3 Using the procedures packaging and bench techniques that are recommended The following practices should be followed to minimize damage to S S static sensitive devices 1 MINIMIZE HANDLING 2 KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE 3 DISCHARGE PERSONAL STATIC BEFORE HANDLING DEVICES USE A HIGH RESIS TANCE GROUNDING WRIST STRAP 4 HANDLE S S DEVICES BY THE BODY static a...

Page 6: ...VE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC FREE WORK STATION PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S S DEVICES 9 HANDLE S S DEVICES ONLY AT A STATIC FREE WORK STATION 10 ONLY ANTI STATIC TYPE SOLDER SUCKERS SHOULD BE USED 11 ONLY GROUNDED TIP SOLDERING IRONS SHOULD BE USED PORTIONS REPRINTED WITH PERMISSION FROM TEKTRONIX INC AND GERNER DYNAMICS POMON...

Page 7: ...ery and re install the battery door Make sure it is securely in place 8 Return the mA Calibrator to its red holster aqv07f eps Figure 1 Replacing the Battery User Replaceable Parts User replaceable parts and accessories are shown in Figure 2 and Table 1 1 2 3 6 5 4 7 aqv10f eps Figure 2 User Replaceable Parts Table 1 User Replaceable Parts Item Part Description Fluke Part No Qty A Non skid foot 88...

Page 8: ...C Amps DCI 6 Refer to Table 4 to verify the readings on the HP 3458A for the following tests No adjustment is necessary for test 1 Test 2 1 Press B twice to select 12 000 mA 2 Verify the readings Test 3 1 Press B twice to select 20 000 mA 2 Press and rotate the knob until 24 000 mA is displayed 3 Verify the readings DC Current Measurement Mode Tests 4 6 1 Press the A key twice to put the mA Calibr...

Page 9: ...and set output to STANDBY 7 Turn the mA Calibrator OFF and disconnect the mA Calibrator from the 5520A 8 The performance verification tests are now complete 9 If the mA Calibrator failed any of these tests calibration adjustment or repair is required Calibration Adjustment Perform the following calibration adjustment procedure if the mA Calibrator fails the performance verification test Calibratio...

Page 10: ...st 6 Disconnect the HP 3458A from the mA Calibrator 7 Connect the test leads from the AUX terminals of the Fluke 5520A to the terminals on the mA Calibrator black to COM and red to 8 The mA Calibrator will display MEASURE and 0 000 mA is displayed On the secondary display the last 3 digits of the A D reading are shown Set the 5520A to output 0 000 mA into the mA Calibrator Press C 9 After a moment...

Reviews: