867B,863
Service Manual
4-14
Duty Cycle Test
4-12.
Rotate the GMM selector to AC Volts ( ). Check that ‘Full Auto’ is selected in the
display above
x
. Select Hz and Duty Cycle. Connect the GMM and the PM5139 in
Configuration 10 (Figure 4-6) and perform the test shown in Table 4-12.
Table 4-12. Duty Cycle Performance Test
Range
Input
Input Source
Lower Limit
Upper Limit
AUTO
50.00% 10 kHz, 0.2V p-p
square
PM 5139
49.8
50.2
FREEZE
DISPLAY
MODE
SAVE
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
1
2
5
3
4
A
COM
V
EXT TRIG
mA/
A
LOCAL
ADDR
x10
OUTPUT
Zo 50
Ω
PM
5139 function generator 0.1 mHz 10MHz
RESET
RANGE
A
mV
V
V
Hi
Z
mA
A
os33f.eps
Figure 4-6. Configuration 10 (PM5139)
Summary of Contents for 863
Page 49: ...Maintenance Disassembly 3 3 5 OFF 1 6 Places 4 5 5 3 3 2 os5f eps Figure 3 1 Disassembly ...
Page 50: ...867B 863 Service Manual 3 6 os6f eps Figure 3 1 Disassembly cont ...
Page 52: ...867B 863 Service Manual 3 8 5 1 6 2 3 4 os30f eps Figure 3 2 Reassembly ...
Page 96: ...6 1 Chapter 6 Schematic Diagrams Title Page 6 1 A1 Main PCA Assembly 6 3 ...
Page 98: ...867B 863 Service Manual 6 4 FLUKE 867 1001 Sheet 2 of 5 Figure 6 1 A1 Main PCA cont ...
Page 99: ...Schematic Diagrams 6 6 5 FLUKE 867 1001 Sheet 3 of 5 Figure 6 1 A1 Main PCA cont ...
Page 100: ...867B 863 Service Manual 6 6 FLUKE 867 1001 Sheet 4 of 5 Figure 6 1 A1 Main PCA cont ...
Page 101: ...Schematic Diagrams 6 6 7 FLUKE 867 1001 Sheet 5 of 5 Figure 6 1 A1 Main PCA cont ...