Model 88 V
Users Manual
60
Replacing the Fuses
Referring to Figure 21, examine or replace the Meter's
fuses as follows:
1.
Turn the rotary switch to OFF and remove the test
leads from the terminals.
2.
Remove the battery door by using a standard-blade
screwdriver to turn the battery door screws one-
quarter turn counterclockwise.
3.
Remove the three Phillips-head screws from the case
bottom and turn the case over.
4.
Gently push up the input terminal-end of the top case
from inside the battery compartment to separate the
two halves of the case.
5.
Remove the fuse by gently prying one end loose, then
sliding the fuse out of its bracket.
6.
Install ONLY specified replacement fuses with the
amperage, voltage, and speed ratings shown in Table
10.
7.
Verify that the rotary switch and the circuit board
switch are in the OFF position.
8.
Replace the case top, ensuring that the gasket is
properly seated and case snaps together above the
LCD.
9.
Reinstall the three screws and the battery door.
Secure the door by turning the screws one-quarter
turn clockwise.
Service and Parts
If the Meter fails, check the battery and fuses. Review this
manual to verify proper use of the Meter.
Replacement parts and accessories are shown in Tables
10 and 11 and Figure 22.