Restarting a Diligence 600 transmitter should not be necessary under normal operation.
However, if a transmitter is having trouble connecting or any problems are experienced
whilst setting up the transmitter, then the transmitter can be restarted by removing one of
the battery cells from the battery compartment
for at least ten seconds
then re-fitting the
battery cell and securely replacing the battery cover.
Should the same issue persist with the transmitter, making setup via the
Setup Mode
impossible, then please contact
or the local distributor for further
assistance.
Should a full reset to factory defaults ever be advised as part of the technical support
process, then the following instructions should be observed:
•
Press and hold both the
WiFi Button
and
Action Button
for
five seconds
to place the
transmitter into
Setup Mode
.
•
Press and hold the
Action Button
continuously for approximately
ten seconds
to initiate
the reset. If successful ‘
CLr
’ will be displayed along with an animation, indicating that the
reset is in process.
•
Once complete the transmitter will revert to a standby mode and will display ‘OFF’
•
The transmitter now needs to be set up again, but this time the server details will also
need to be set. Please refer to
for full details on accessing the
Advanced
Configuration
and configuring the
Server Settings
.
34
Copyright © 2022 Comark Instruments