DATAPAQ TP3
Using Insight Software
45
Multiple Runs Rather than having to download data at the end of every
profile run (the default behavior), the logger can collect data for up to 10
runs before downloading (see also p. 49, p. 51). To enable this multiple-
run mode, first click Advanced in the Reset dialog, then click the ‘Multiple
Runs’ button which appears.
In single-run mode
(
the default, if ‘Multiple Runs’ is not selected
)
:
○
Data must be downloaded after every run.
○
When starting the next profile run, the data held in the logger will be
deleted automatically (this is prevented if the data has not yet been
downloaded), so there is no need to reset the logger unless you wish
to change the data-collection options. (See p. 47 for starting a new
run using previous reset options.)
○
Any trigger mode (see p. 43) can be selected.
○
Hardwired, Bluetooth and radio telemetry can all be used.
In multiple-run mode:
○
The logger can collect data from up to 10 profile runs before it needs
to be downloaded (p. 48).
○
After 10 profile runs have been performed, the logger must be reset
using Insight in order to delete data from the logger and empty it for
further runs.
○
The memory calculator shows the total time available for all runs until
the logger is reset.
○
Individual profile runs can be downloaded from the logger at any time,
but this does not free up any space in the logger’s memory until the
whole memory is emptied by performing a reset using Insight.
○
Each multiple run will be performed using the same data-collection
options, until the logger is reset.
○
To
start data-collection
for each run, only the start button and
temperature
trigger modes
can be selected (see p. 43).
○
Stop data-collection
at the end of each run by using the stop button. If the
stop button has been disabled (see p. 53), instead set a
stop trigger
to
stop collection according to rising/falling temperature or specified
duration (see p. 44). Pressing the stop button when paused between
runs will always terminate multiple-run mode.
○
Hardwired and Bluetooth telemetry cannot be used (but radio
telemetry can be used).
Multiple Events By default, the same probe selections, sample interval and
telemetry options are used for the whole of a profile run – but multiple-event
mode allows you to specify different settings within different stages of the
run, each new stage being triggered by a specified combination of time and/or
temperature. To enable this multiple-event mode, first click Advanced in the
Reset dialog, then click the ‘Multiple Events’ button which appears; this
further expands the dialog and activates a default minimum of two events.
Select the event-number you wish to modify, and make selections.
Summary of Contents for Datapaq TP3
Page 2: ......
Page 8: ......
Page 10: ...10 Introduction DATAPAQ TP3 ...