Installation
5.2.3
Accessory Cables and Terminal Block
As accessories for the Endurance 1R devices there are two different communication cables and a
specific terminal block available. Both sensor cables can be ordered in several cable lengths and two
different ambient temperature ratings.
The sensor head is rated NEMA-4 (IEC 529, IP65).
Endcap must be securely installed to maintain proper sealing.
To prevent possible electrical shock, fire, or personal injury make sure that the
sensor is grounded before use.
5.2.3.1
M16 12-Conductor shielded cable
The 12-conductor shielded connecting cable is used to wire all the fundamental inputs and outputs
like RS485 interface, trigger input, relay contact, current input, current output and 24V power supply
wires to the Endurance 1R sensor. The cable is equipped with an IP67 rated M16 12-socket DIN
connector at one end and colored wires with cable end sleeves at the counter side.
See below the colored wire to signal assignments, which are identical to the specific terminal block
labeling. For more cable details, see section 9.5 of this manual
Figure 15: M16 12-Conductor shielded cable with colored wire/signal assignments
If you cut the cable to shorten it, notice that both sets of twisted-pair wires have drain
wires inside their insulation. These drain wires (and the white wire that is not part of
the twisted pair) must be connected to the terminal labeled CLEAR or SHIELD.