Endurance® Series
Users Manual
86
Analog input mode
INM
n
0 = 0-20mA, 4 = 4-20mA
4
IP Address
IP
nnn.nnn.nnn.nnn
√
√
0.0.0.1 - 255.255.255.255 192.168.42.132
Switch panel lock
J
X
√
√
L = Locked
U = Unlocked
Unlocked
Relay alarm output
control
K
n
√
√
0 = Permanently Open
2
1 = Permanently Closed
2 = Normally Open
3 = Normally Closed
Bottom of mA
temperature range
L
n.n
– nnnn.n
√
√
√
0.0
– 9999.0 (°C or °F)
Lower end of
sensor range
Mode
–ER series
M
n
√
√
√
1 = 1 - color
2
2 = 2 - color
MAC Hardware Address
MAC
nnnnnnnnnnnn
√
e.g. 001d8d2aaa01
Set at factory
calibration
Target temp:1C narrow
N
n.n - nnnn.n
√
√
Net Mask
NM
nnn.nnn.nnn.nnn
√
√
0.0.0.1 - 255.255.255.255 255.255.255.0
Output current
O
nn
√
√
√
00 = controlled by unit
00
02 = under range
21 = over range
00
– 20 = current in mA
Output format
OIF
0 = temperature value
format is float with one
decimal place
1 = temperature value
format is integer with four
digits
0
Peak hold time (4)
P
n.n - nnn.n
√
√
√
0.0
– 300.0 sec (300 s =
)
0.0
IP Portaddress
PORT n - nnnnn
√
√
1 - 65535
6363
Wide Power
Q
n.nnnnnnn
√
√
Narrow power
R
n.nnnnnnn
√
√
Video relative reticle
diameter
RC
n.n
– nn.nn
√
Video relative reticle X-
position
RX
n.n
– nn.nn
√
Video relative reticle Y-
position
RY
n.n
– nn.nn
√
Slope
S
n.nnn
√
√
√
0.850
– 1.150
1.000
Alarm source
SAS
√
√
0 = object temperature
1 = internal temperature
0
Slope source
SS
X
√
I or E
I
Set target temperature
STT
n.n
– nnnn.n
√
√
0.0
– 9999.0 (5)
Set at factory
calibration
Target temp: 2C
T
n.n - nnnn.n
√
√
Terminator resistor
TR
n
√
√
0 = OFF, 1 = ON
0 = OFF
TCP/IP time out interval
TTI
n - nnn
√
√
0 =
, 1
– 240 sec
0
Temperature units
(scale)
U
X
√
√
√
C or F
non-US: C
Poll/burst mode
V
X
√
√
B = Burst , P = Polled
P = Polled
Target temp:1C wide
W
n.n - nnnn.n
√
√
(5)
Web server ON/OFF
WS
n
√
√
0 = OFF, 1 = ON
0 = OFF
Burst string contents (3)
X$
√
Multidrop address
XA
nnn
√
√
√
000 to 032
000
Low temperature limit
XB
n.n - nnnn.n
√
0.0
–9999.0 (5)
Set at factory
calibration
Deadband (7)
XD
nn
√
√
01
– 55 in °C / 01 – 99 in
°F
02
Restore factory defaults
XF
√
Transmissivity
XG
n.nn
√
√
√
0.10
– 1.10
1.00
www.
.com
information@itm.com
1.800.561.8187