Description
Command
Example
Response
Format
Response
Example
Range
Set autolog mode on/off .
Note:
Exit Autolog menu
to use this command. After
alog =ON is sent, the buttons
are disabled until the command
alog=OFF is sent.
alog=[ON/OFF]
alog=ON
ON or OFF
Clear autolog memory
aclr
aclr
Initiate download of autologged
data.
Note: Read Section
11.2 on downloading
autologged data before at-
tempting to use this
command.
daut
daut
Read autolog interval
aint
aint
int: 99 {secs or min}
int: 60 {secs or min}
Set autolog interval
Note: See Section 7.4.2
for specifics on log interval.
aint=n
aint=30
1, 5, 10, 10,
15, 30, 60,
120, 300, 600,
900, 1800,
3600
Read delta(x) measurement
dx
dx
dx: 999.999
{C, O, F, R, or K}
dx: -0.118 C
Set delta(x) reference to pres-
ent measurement
sd[x]
sd
Read unit
u[nit]
u
u: {C, O, F, R, or K}
u: C
Set unit
u[nit]=C/O/F/K/R
u=C
C, O, F, K, R
Read serial sample rate
sa[mple]
sa
sa: 99999
sa: 10
Set serial sample rate
sa[mple]=
n
sa=60
0 to 10,000
Read resolution digits
re[solution]
res
res: 9
res: 3
Set resolution digits
re[solution]=
n
res=0
1 to 3
Read time
ti[me]
time
ti: hh:mm:ss
ti: 16:23:45
Set time
Note: You must set
hours, minutes, and seconds
for the time to be set
correctly.
ti[me]=hh:mm:ss
ti=08:15:00
00:00:00 to
23:59:59
Read identification information
*i[dn?]
*idn?
<manufacturer>,
<model>,
<serial number>,
<firmware version>
Hart,
1522
95001,
1.24
Read model and firmware
version
*v[ersion]
*ver
ver.<model>,
<firmware version>
ver.1522,
1.24
Legend:
[] Optional Command data
{} Returns either information
49
8 Communications Interface
Communication Command List