Technical Data
14
ScanIR3
2.3
Electrical Specifications
Communications
Ethernet
TCP/IP protocol 10/100 MBit/s, electrically isolated
auto-negotiation
Inputs
Trigger
+ 5 to 24 VDC pulse, high/low active
Functional Input
max. + 5 VDC
function depends on the specific linescanner system
Signal Processing
Stand-alone unit
Max, Min, Average, Peak/Valley Hold, Alarm setpoints
further signal processing function through software
Power
AC Power
100 to 240 VAC, 50/60 Hz, 0.6 A
Altitude:
2000 m (6562 ft)
Terminal Wiring
AC Power
1.5 mm² (AWG 16) to 2.5 mm² (AWG 14)
I/O Modules
1.5 mm² (AWG 16)
2.4
General Specifications
Mechanical Scanning System
40,000 h (about 4.5 years) meantime between failures (MTBF)
for the mechanical scanning system at 48 Hz scan rate
Environment Rating
IP65 (NEMA 4) in accordance to IEC 60529
Ambient Temperatures
All models
0 to 50°C (32 to 122°F) – without cooling
max. 180°C (356°F) – with integrated water cooling (standard)
S335
down to –40°C (–40°F) with internal heating element
optional for all other scanner models
Storage Temperatures
-25 to 65°C (-13 to 149°F)
Water Cooling: Pressure
max.
15 bar (218 PSI)
Water Cooling: Flow rate
typ. 1 l / min to 2 l / min (0.26 to 0.52 gallons / min)
Air Purge Collar: Pressure
max. 3 bar (43 psig)
Air Purge Collar: Flow rate
typ. 100 l/min to 200 l/min (3.53 to 7.06 cfm)
Warm-up Time
30 min.
Relative Humidity
10 to 90%, non-condensing,
for operating and non-operating temperature range
Vibration
IEC 60068-2-6, 3 axes, 10 to 150 Hz, operating 2 g above 20 Hz
Shock
IEC 60068-2-27, 3 axes, operating: 5 g at 11 ms, 15 g at 6 ms
1
at still air (e.g. by using a protective housing)