Installation
ScanIR3
33
5.7
Air purge collar
The air purge system produces a laminar air flow that protects the linescanner window from dust,
moisture, and vapor. The air flows from the couplings through the walls of the housing and through
side slits near the scanner’s window. The air flow should be between 100 l/min (3.53 cfm) and 200
l/min (7.06 cfm) through each side, which corresponds to a pressure between 0.5 bar (7.25 psig) and
3.0 bar (43 psig) when using the supplied metric fittings. The inside diameter for the air hose should
be 4 mm (0.16 in.). The hose couplings for the air purge system are connected to the housing through
ISO 228 G 1/8” stainless steel fittings.
Use only clean or “instrument grade” air (free from oil contaminants). Do not use
cooled air. This could lead to condensation on the linescanners window!
5.8
Laser
The built-in laser sighting function allows fast and precise aiming at small or rapidly moving targets,
or targets passing at irregular intervals. The laser is specially aligned with the scan line of the
linescanner. A small, bright red laser line shows you the center of the scanned line, not the size of the
spots being measured.
The laser can be turned on and off via the scanner runtime software, see menu <Scanner> <Switch
Laser on/off>.
To preserve laser longevity, the laser automatically turns off after 10 minutes of
constant use!
Technical Data:
Laser type
Laser diode
Wavelength
635 nm, red
Internal temperature
automatic switch off at < 5°C (41°F) or > 50°C (122 °F)
Output power
1 mW
Laser class
2
Horizontal angle
70°, non-symmetrically
W
ARNING
!
Avoid exposure to laser light! Eye damage can
result. Use extreme caution when operating!
Never point at another person!