™/S, MOLBLOC TERMINAL OPERATION AND MAINTENANCE MANUAL
© 1995 - 2010 Fluke Corporation, DH Instruments Division
Page 94
OPERATION
To set the unit of measure in which displays pressure values, press
[SETUP]
and
select
<8presU>
.
The display is:
Pressure unit type:
1SI 2other 3user
Select the unit type desired, then select the unit desired. After the unit selection, operation
returns to the run screen with the selected pressure unit of measure active.
The pressure units of measure available are listed in Table 33.
Table 33.
Pressure Units of Measure Available
<1SI>
<2Other>
<3User>*
<1Pa>
<2kPa>
<3mPa>
<4mbar>
<5bar>
<6mmHg>
<7mmWa>
<1psi>
<2psf>
<3inHg>
<4inWa>
<5kcm2>
<1user>
*3User: User defined unit.
Note
•
The “user” unit is defined in terms of user units/Pa when the user unit
is selected.
•
See Section 7.1.1 for definition of the pressure unit conversions used
by .
3.5.9
<9TEMPU>
PURPOSE
To select the unit of measure in which displays temperature values.
OPERATION
To set the unit of measure in which displays temperature values, press
[SETUP]
and select
<9tempU>
.
The display is:
Temperature unit:
1celcius 2fahrenheit
Select the desired unit. After the unit selection, operation returns to the run screen with the
selected temperature unit active.
Note
See Section 7.1.2 for definition of the temperature unit conversions used
by .