3. OPERATION
Page 97
© 1995 - 2010 Fluke Corporation, DH Instruments Division
•
Pressure unit of measure to kPa (see Section 3.5.8).
•
User pressure unit coefficient to 1.00/Pa (see Section 3.5.8).
•
Temperature unit of measure to °C (see Section 3.5.9).
3.6.1.3
<3MFC>
PURPOSE/OPERATION
To access Reset - MFC, press
[SPECIAL]
and select
<1reset>
,
<3MFC>
.
Reset - MFC clears and sets to default all functions associated with the optional
analog MFC control. These include:
•
Set MFC channel to 0 (inactive) (see Section 3.4.8).
•
Set regulation mode to OFF (see Section 3.5.6).
•
Set MFC profile to 1 (see Section 3.4.8.2).
•
Clear/delete user defined MFC profiles (see Section 3.5.5).
3.6.1.4
<4cal>
PURPOSE/OPERATION
Caution
Use special caution with this reset as critical calibration data
may be altered.
To access Reset - Cal, press
[SPECIAL]
and select
<1reset>
,
<4cal>
.
Reset - Cal clears and sets to default the user calibration coefficients for
Reference Pressure Transducers (RPTs) (see Section 5.2). This
includes:
•
Upstream and downstream absolute RPTs:
Adder
0
Multiplier:
1
Calibration Date: 19980101
Note
Reset - Cal has NO effect on the reference resistance values
used to calibrate s internal ohmic measurement
system (see Section 5.4) or on the calibration coefficients for the
optional MFC control function (see Section 3.4.8).
3.6.1.5
<5all>
PURPOSE/OPERATION
To return to the original, as delivered factory condition. Performs the
SETS, UNITS, CAL and MFC RESET functions and resets all other settable
values to defaults. This includes communications port settings.
To access Reset - All, press
[SPECIAL]
and select
<1reset>
,
<5all>
.
Caution
Use special caution with this reset as critical calibration data
may be altered.