3.12.1.14 Delete A Mass Loading Bell
PURPOSE
To delete an existing mass loading bell file entirely.
OPERATION
Caution
Once a mass loading bell file has been deleted it cannot be
recovered. To recreate it, the add mass loading bell function
must be used (see Section 3.12.1.11) which requires entering all
the mass loading bell information. Before deleting a mass
loading bell, consider editing it (see Section 3.12.1.12).
To delete a mass loading bell press
[SPECIAL]
,
<1PC/MS>
and select
<3mass
bell>
,
<4delete>.
Confirmation to delete is requested. Select
<2yes>
to delete the mass loading bell.
Press
<1no>
to NOT delete.
Note
The active mass bell (mass bell that is currently selected)
cannot be deleted. Before attempting to delete a mass bell, be
sure it is not currently selected.
3.12.1.15 Select a Mass Loading Bell
PURPOSE
To select the mass loading bell that is active and used by PG9000 in its mass
loading and calculation of the reference pressure.
OPERATION
To select the active mass loading bell set press
[SPECIAL]
,
<1PC/MS>
and
select
<3mass bell>
,
<5select>
.
3.12.2 <2presU>
PURPOSE
To customize the selection of pressure units that are available for selection from the UNIT
function key (see Section 3.10.3).
PRINCIPLE & OPERATION
When the
[UNIT]
key is pressed there are six units of measure. These six are displayed as a
convenience to quickly access and change unit of measure. To edit the display, the
[SPECIAL],<2presU>
is used to select the desired unit and the place holder where it is to be
displayed. See Section 3.10.3.1.
3.12.3 <3head>
PURPOSE
To edit the DUT head configuration; to edit the ATM head height; to turn the PISTON head
correction ON and OFF.
PRINCIPLE
PG9000 supports three different fluid head correction functions.
1. There is a correction to consider the difference in height between the PG9000 reference
level and a device under test (see Section 3.10.7, PRINCIPLE). This head correction is
referred to as DUT head. The head height can be adjusted by pressing the
[HEAD]
key.
The head unit of measure and fluid are adjusted by pressing
[SPECIAL]
and selecting
<3head>
,
<1fluid>
and
<2unit>
.
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...