PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
80
OPERATION
Note
For details on pressure-to-mass and mass-to-pressure operation, see
Section 3.10.11.
To access the P
OR
M function, press
[P
OR
M]
,
the display is:
Select entry mode:
1pressure 2mass
Selecting
<1pressure>
activates
pressure-to-mass
mode and returns to the previous run screen.
Selecting
<2mass>
activates
mass-to-pressure
mode and returns to the previous run screen.
3.10.13 [
] and [ ] + [
←
]
PURPOSE
To manually activate motorized piston rotation or braking function.
PRINCIPLE
Under normal operations, it is recommend that the motorized rotation should be enabled to
automatically engage and disengage the rotation of the PG9000 piston.
Motorized rotation can be set to operate automatically to maintain the piston within
Ready
condition rotation rate limits when the piston is floating and stop rotation before changing the
pressure or mass load (see Section 3.10.8). Motorized rotation can also be engaged
manually to accelerate or brake the piston rotation at any time under direct operator control.
[
]
is used for momentary acceleration of piston rotation. Pressing and holding
[
]
followed by
[
←
]
is used to start a function that stops piston rotation.
OPERATION
To momentarily engage the motorized piston rotation system and accelerate the piston, press
and hold
[
]
from any run screen. The motorized rotation system engages and remains
engaged until maximum rotation rate has been achieved or until the key is released.
The display is a modified version of the 1
st
system run screen to indicate the piston rotation is
being accelerated while showing rotation rate and position:
1.
Measured piston rotation rate.
2.
Current rate of piston vertical displacement.
3.
Measured piston position.
12 rpm ACCELERATING
+ 2.05 mm 0.1/min
To start the piston braking function press and hold
[
]
, then press
[
←
]
. The motorized
rotation system engages and stays engaged until the piston rotation is stopped. Once the
braking function starts the keys may be released and the function will complete unless
[ESC]
is pressed.
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...