Edit mass loading bell information (see Section 3.12.1.12).
•
View the active mass loading bell (see Section 3.12.1.13).
•
Delete a mass loading bell (see Section 3.12.1.14).
•
Select a mass loading bell (see Section 3.12.1.15).
3.12.1.1 Create a Piston-Cylinder Module
PURPOSE
To create a new piston-cylinder module that will be available for selection from
the
[P-C]
function key. Use this function when a new piston-cylinder module has
been acquired.
Note
If the piston-cylinder module and PG9000 Platform were
delivered together, the piston cylinder metrological data has
already been entered at the factory. Before creating a new
piston-cylinder module, press [P-C] to check if it already exists
(see Section 3.10.2).
PRINCIPLE
The PG9000 add and edit piston-cylinder module functions allow piston-cylinder
metrological data to be defined and stored. This data will be used by PG9000 in
calculating the defined pressure, piston-cylinder range, pressure-to-mass and
mass-to-pressure values when the piston-cylinder module is made active. The
required data can be found in the calibration report for the piston-cylinder
module.
Setting up or editing a piston-cylinder module requires specifying, in sequential
order, the following:
Serial number (S/N)
[nnnn]
.
Effective area (Ae) at 0 kPag pressure and 20 °C
[mm2]
.
Piston temperature coefficient
[linear thermal expansion/°C]
.
Cylinder temperature coefficient
[linear thermal expansion/°C]
.
Effective area pressure coefficient
[change in Ae/MPa]
.
Piston-cylinder surface tension effect
[N/m]
.
Reference level offset
[mm]
.
Piston assembly mass (true mass)
[kg]
.
Piston assembly average density
[kg/m
3
]
.
Rotation rate ready limits
[rpm]
.
k(P)
[kg/mm/min]
.
Calibration report number
[nnnnnnnnnn]
.
Calibration report date
[yyyymmdd]
.
OPERATION
Note
•
PG9000 can store up to 18 piston-cylinder module (PC) files.
When <3add> is selected and all the files have already been
used, a warning is displayed -- a file must be deleted before
a file can be added.
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...