PK-PPC-BG-CVU Dual Volume Unit and 

Interconnections Kit 

P/N 

 

3070389

 

 

 

Interconnections Kit, P/N 3070389, includes: 

DESCRIPTION 

QTY 

FIGURE 

CALLOUT 

PART NO. 

Dual Volume Unit (DVU) 

N/A 

3119872 

Tubing,  PFA,  1/4  in.  x  160 
cm 

3232378 

Adaptor,  1/4  in.  tube  x  O-
seal 1/8 in. NPT M 

3140484 

Adaptor,  1/4  in.  tube  x  1/4 
in. NPT M 

3134427 

Union, 4P 

3135104 

Tee, 1/4 in. tube 

3134453 

Adaptor,  1/4  in.  tube  x  1/4 
in. NPT F 

3134430 

Set, 6 each, Rubber Shoes 

N/A 

3125710 

 

To Install the Dual Volume Unit (DVU) 

   Identify a location for the DVU™ that minimizes 

the  length  of  tubing  between  the  DVU 
connections  and  the  PPC  BG15K™  TEST(+)  and 
TEST(-) ports.  The PPC BG15K and the DVU are 
designed to stack one on top of the other.  The 
DVU  may  be  rack  mounted  using  the  rack 
mount  kit  (P/N  3069019).    Recommended 
configurations  are  for  the  PPC  and  DVU  to  be 
stacked directly one on top of the other or with 
their  rear  panels  face  to  face.  For  PPC4  attach 
rubber shoe set to the legs of the DVU.  

To Connect the DVU and PPC BG15K 

 Note

 

You will be making and connecting 
two identical assemblies, one for 
the TEST(+) port and one for the 
TEST(-) port. 

   Install  a  union  (4)  into  a  1/4  in.  tube  x  1/4  in. 

NPT  M  adaptor  (3)  (do  this  twice).    Screw  the 
1/4 in. NPT M end of the adaptors into the two 
ports of the DVU using Teflon

®

 tape as a sealer. 

 

Figure 1:  

Interconnections Kit

 

  Install  a  1/4  in.  tube  tee  (5)  onto  each  of  the 

union/adaptor  assemblies  that  were  just 
installed  into  the  DVU  ports  in  Step  .    Orient 
the  tee  considering  the  connection  to  the  PPC 
BG15K  (schematic  above  is  for  PPC  stacked  on 
top of the DVU).

 

  

Install the 1/4 in. tube x O-seal 1/8 in. NPT M (2) 
into the PPC BG15K TEST(+) and TEST(-) port. 

  Cut lengths of PFA tubing as needed to connect 

the 1/4 in. tube tee to the adaptors on the PPC 
TEST ports.  Take care that the cut is at a right 
angle  to  the  tube  and  only  cut  the  length 
needed.  

   Optional for a connection “For DUT” of 1/4 in. NPT 

M rather than 1/4 in. tube: 

•  Cut  a  small  length  of  PFA  tubing  to  use  as  

a union between the 1/4 in. tube tee (5) and  
the  1/4  in.  tube  x  1/4  in.  NPT  F  adaptor  (6) 
(see schematic above). 