Local Operation
Menu
4
4-9
Menu | Setup — Units
In addition to the standard units of measure provided by the PPI, four user-defined units
are available. To create a user-defined unit, the user enters a
name
that is one to ten
characters long and a
conversion factor
that is a multiple of kiloPascals (kPa).
For example, using the information from Table 2-1, the conversion factor for millitorr or
one micron of mercury at 0
°
C is calculated as follows.
kPa
psi
x
psi
C
mmHg
x
C
mmHg
Torr
x
Torr
mTorr
x
kPa
mTorr
1
1450377
.
0
0193377
.
0
0
1
0
1
1
1
1000
°
°
=
thus the conversion factor simplifies to
6180
.
7500
x
kPa
mTorr
=
4-11.bmp
Figure 4-11. Menu | Setup | Units - Menu
1.
The pressure units are defined from the Units | Define | Menu. From the Main Menu
(press PREVIOUS until the Main Menu appears) press Menu [F6] | Setup [F2] |
Units [F3].
2.
Use the rotary knob to highlight the desired user-defined unit, and then select the
Edit Name [F1] function key.
3.
The following sequence is used to change the name of the selected unit.
4-12.bmp
Figure 4-12. Menu | Setup | Units | Edit Name - Menu
Summary of Contents for RUSKA 7050