RUSKA 7050, 7050i & 7050LP
Users Manual
5-2
RS-232
The RS-232 interface supports standard serial operation from a computer to a single PPI.
RS-232 supports the IEEE-488.2 and SCPI commands. The PPI allows the following port
setups:
Baud
Rate
1200, 2400, 9600, or 19200
Data
Bits
7 or 8
Parity
Even,
Odd,
or
None
Stop
Bits
1 or 2
Handshaking
XON/XOFF
The RS-232 connection is a DB-9P connector found on the back panel of the PPI. It is
located on the processor board. The following pins are used; all other pins are reserved.
Pin
#
Direction Signal
2 In
RXD
Receive
Data
3 Out
TXD
Transmit
Data
5 ——
GND
Ground
7 Out
RTS
Request
to
Send
Remote/Local Operation
In Local mode, the PPI is operated manually through the front panel. Chapter 4 covers
local operation. The PPI always powers up in the Local mode. In Remote mode, the PPI
is operated by a computer connected to an interface. Most functions that can be
performed in Local mode can also be performed remotely.
Remote mode does not automatically disable local operation. The remote interface may
be active while local operations are being done. In cases where full remote control is
required, the following methods may be used.
1.
Issue a Local Lockout (LLO) interface message via the IEEE-488 interface. The PPI
will disable the local keyboard until the Go To Local (GTL) interface message is
received or the REN (Remote Enable) line is unasserted. This method cannot be used
on the serial interface.
2.
Issue the SCPI command “SYSTEM:KLOCK” on to lock the local keyboard. The
PPI will disable the local keyboard until the command “SYSTEM:KLOCK OFF” is
received.
3.
Issue the SCPI command “DISPLAY:ENABLE OFF” or
“DISPLAY:TEXT <string>.” These commands will disable the local display in
addition to locking the keyboard. The command “DISPLAY:ENABLE ON” will
restore the local display and keyboard operation.
Local operation may also be restored by turning the PPI off and back on.
Configuration
The remote interface must be configured before it is connected. The remote interface is
configured using the local interface. The parameters needed vary with the interface used.
IEEE
-
488 Address,
Protocol
RS
-
232
Baud Rate, Data Bits, Parity, Stop Bits
Summary of Contents for RUSKA 7050
Page 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Page 4: ......
Page 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Page 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Page 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Page 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Page 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...