Remote Operation
Druck DPI-510 Emulation Command Summary
5
5-9
R1,R2
Lock
keyboard
S0,S1,S2,S3
Set pre-programmed units
T0
Turn off Tare mode
T1
Turn on tare mode to value set to B (B must be set first)
U1 to U23
Set S4 units (if S4 is selected this will change current units
immediately)
V<value>
Set slew rate (sets low overshoot mode)
W<value>
Set In Limit wait time
/0 to /11
Set set-point as fraction of P value
*
0 to *11
Set pre-programmed set-point
@0
Disable
error
reporting
@1
Enable
error
reporting
Response Formats
N0
<Pressure or Setpoint><REM|LOC>RnSnDn[@nn]
N1
<Pressure or Setpoint>[@nn]
N2 <REM|LOC>RnSnDnCnInFn
N3
<In Limit Status>[
@
nn]
N4
@
nEnJnVnnnnUaaaa
Note
Secondary addressing is not supported.
Reset Conditions
N0 D0 F0 I0 @1 R0 S0 W002
Keyboard Entry
MENU/SETUP/REMOTE/510EMUL
Termination:
CR/LF/EOI, CR/EIO, LF/EOI, or EOI
Zero Delay:
Delay for zero stabilization.
F1,F2,F3 Rates Specify rates used by J0, J1, J2
MENU/SETUP/REMOTE/510EMUL/F1UNIT
MENU/SETUP/REMOTE/510EMUL/F2UNIT
MENU/SETUP/REMOTE/510EMUL/F3UNIT
Select units for S0, S1, S2
MENU/SETUP/REMOTE/510EMUL/RATIOS
Specify ratios used by /0 to /11
MENU/SETUP/REMOTE/510EMUL/PRESETS
Specify presets used by *0 to *11
Summary of Contents for RUSKA 7050
Page 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Page 4: ......
Page 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Page 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Page 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Page 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Page 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...