RUSKA 7050, 7050i & 7050LP
Users Manual
5-10
Notes
Format
Upper and lower case are equivalent. Spaces and equal signs are ignored before numeric
values. Spaces, commas, semi-colons, and colons are ignored before and after commands.
Zeroing
The RUSKA sensor takes longer to zero. This time can be set from the front panel. The
default is 5 seconds but this may need to be increased especially for absolute units. When
the zero command is received the zeroing valve is opened, the system waits the zero
delay time and then the sensor is zeroed.
Rate
If the rate is positive it will be used at the maximum rate with the overshoot set to low. If
the rate is zero then the maximum rate will be used with the overshoot set to low. If the
rate is negative the maximum rate will be used with the overshoot set to normal.
Units
Unit Number
Unit Name
RUSKA display units
1
Pa
kPa
2 kPa
kPa
3
Mpa
kPa
4
mbar
bar
5 bar
bar
6 kg/cm2
kg/cm
2
7 kg/m2
kg/cm
2
8 mmHg
mmHg
0
o
C
9
cmHg
cmHg 0
o
C
10
mHg
cmHg 0
o
C
11
mmH2O
cmH
2
O 4
o
C
12
cmH2O
cmH
2
O 4
o
C
13
mH2O
cmH
2
O 4
o
C
14
torr
mmHg 0
o
C
15
atm
bar
16
psi
psi
17
lb/ft2
psi
18
inHg
inHg 0
o
C
19
"H2O04 inH
2
O 4
o
C
20
'H2O04
InH
2
O 4
o
C
21
SPEC'L
User
Defined
1
22
"H2O20 inH
2
O 20
o
C
23
'H2O20
inH
2
O 20
o
C
Summary of Contents for RUSKA 7050
Page 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Page 4: ......
Page 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Page 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Page 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Page 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Page 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...