6-10
Zeroing
The zeroing procedure is performed to correct for system zero shift and does not require a
full calibration. The most important requirement for performing a valid zeroing procedure
is to guarantee that there is not a pressure differential between the sensor’s test port and
case reference.
If during the zeroing procedure, the message “Mechanical Zeroing Needed” appears, the
sensor photocell may need to be zeroed. See Chapter 6, Sensor Photocell Zeroing, for
more information.
The zeroing screen presents several pieces of information.
The screen will display the current status of the pressure sensor Stability and
Temperature. If any of the above are unstable, then the system will delay until stability is
achieved. Pressing OK [F6] will bypass this wait period.
Note
By-passing this wait period can have a negative effect on the zeroing
procedure.
Gauge and Vacuum (Negative Gauge) Instruments
1.
Verify that the Reference Port is open to atmosphere.
2.
Enter the Calibration screen by selecting MENU | CALIBRATION.
3.
Select Zero [F1]. Do NOT press the Calibrate button.
6-4.bmp
Figure 6-4. Zeroing Menu
4.
Do not disturb the instrument while zeroing is in process.
5.
Wait for the zeroing procedure to finish.
6.
Press PREVIOUS to return to the Main Menu.
Summary of Contents for RUSKA 7050
Page 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Page 4: ......
Page 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Page 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Page 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Page 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Page 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...