RUSKA 7050, 7050i & 7050LP
Users Manual
A-2
characteristics of their device. Some manufacturers identify that their stability
specification is proportional with time. Therefore, if the calibration interval were reduced
in half, the magnitude of the stability specification would also be halved. This can be a
powerful tool when you are trying to improve the measurement performance of a
standard. By reducing the calibration interval, the expanded uncertainty would also
reduce. On the other hand, some manufacturers do not claim that their stability
specification is proportional with time. This would be the case for instance if the
instrument naturally drifted in a sinusoidal fashion. This would suggest that the sensor
could drift to its maximum stability limit at any time and therefore, reducing the
calibration interval would not improve the expanded uncertainty of the device.
Uncertainty of the Standard
Uncertainty of the Standard used to calibrate the transfer standard. This is the expanded
uncertainty of the calibration standard that was used by the manufacturer to calibrate the
digital transfer standard. This should be the expanded uncertainty of the calibration
standard and include all sources of uncertainty that would influence the calibration
standard including the uncertainty from the National Standards Laboratory that the
standard is traceable.
It should also be noted that when the instrument is re-calibrated, the uncertainty of the
device is influenced by the uncertainty of the calibration standard that will be used to
perform the re-calibration. Therefore, the uncertainty analysis should be evaluated
following each re-calibration. If the instrument is re-calibrated using a different
calibration service provider than the manufacturer, the uncertainty of the standard that the
calibration service provider used to perform the calibration would need to be substituted
for the manufacturer’s calibration uncertainty that was used in the original uncertainty
analysis.
Environmental or Installation Influences
Environmental or Installation Influences that could cause errors in the transfer standard
includes influences such as ambient temperature, line pressure, head pressure, time
response, and controller effects. (It may include other influences that are very specific to
one manufacturer’s instrument.) It is recommended that the intended application is
reviewed to assure that the environmental does not impact the instruments performance,
or that the impact from the environment is accounted for in the uncertainty analysis. For
instance, if an instrument has a 0.001%FS per degree Celsius temperature effect from a
calibrated temperature of 20 degrees Celsius, and the instrument is to be used in an
environment where the temperature will vary from 15 to 25 degree Celsius, then a
+/- 0.005% of full scale uncertainty should be included in the uncertainty analysis for
ambient temperature effects.
Combining the Uncertainty Components
Combining the Uncertainty Components into an Expanded Uncertainty Statement means
that once all of the uncertainty components are identified, they can be combined into an
overall Expanded Uncertainty Statement for the instrument. The vast majority of the
measurement community has adopted a statistical approach to uncertainty analysis such
as the International Organization for Standards (ISO) Guild to the Expression of
Uncertainty in Measurement. This statistical approach combines all of the uncertainty
components in quadrature (RSS).
Summary of Contents for RUSKA 7050
Page 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Page 4: ......
Page 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Page 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Page 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Page 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Page 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...