ThermoJacket
Users Manual, Rev. 1.1, Jan 2019
12
2 Technical Data
2.1 General Specifications
Air purge flow
35 – 48 l / min (1.24 to 1.7 foot³ / min.)
Ambient temperatures
water cooling
315°C (600°F)
air cooling
115°C (240°F)
Coolant pressure
(min./max.)
water cooling
2.7 bar (40 psi) to 8.6 bar (125 psi)
air cooling
5.5 bar (80 psi) to 8.3 bar (120 psi)
filtered or instrument-clean air is required
Table 2-1: Ambient Temperatures versus Coolant Flow
Max. Ambient Temperature
Water Cooling Flow
Air Cooling Flow
93°C (200°F)
121°C (250°F)
149°C (300°F)
232°C (450°F)
315°C (600°F)
0.3 l / min (0.01 foot³ / min)
0.6 l / min (0.02 foot³ / min)
1.0 l / min (0.035 foot³ / min)
1.3 l / min (0.046 foot³ / min)
2.0 l / min (0.07 foot³ / min)
95 l / min (3.3 foot³ / min)
110 l / min (3.9 foot³ / min)
Water/air temperature of 20°C (68°F) at inlet assumed.