3.3.2 Reassembly of the Tube Fittings
You may disassemble and reassemble a Swagelok tube fitting as often as required.
1. Insert tubing with pre-swaged ferrules into the fitting
body until the front ferrule seats.
2. Rotate the nut with a wrench to the previously pulled-up
position. At this point, a significant increase in
resistance will be encountered.
3. Tighten slightly with a wrench.
3.4 Air Cooling
The H
2
O ports are also designed to use air for cooling, for further information see section 3.3
page 17.
To control air flow, use the air flow regulator accessory, see section 4.9
3.5 Air Purging
Connect the air line to the female 1/4” NPT air-in port. Control the air flow with an air flow regulator. Air flow for
purging should be approximately 46 l / min (1.6 foot³ / min) and filtered through a pressure regulator with an integral
filter to prevent oil and contaminant build-up on the sensor lens. Connect the air flow regulator between the
ThermoJacket and the air pressure regulator. To ensure leak-free connections, use Teflon tape or equivalent pipe
sealant.