3.3. Measurement and Optical Specifications
Measurement Specifications
Device Model
Parameter
TV43 & TV46
Measuring range
Standard: -10°C to 1200°C (14°F to 2192°F)
Accuracy
± 2°C or ± 2% of reading (whichever is greater)
Infrared resolution (number of pixels)
TV43: 320 x 240 (76800)
TV46: 640 x 480 (307200)
Spectral range
8 – 14 µm
Emissivity correction
0.10 to 1.00
Field of View (aperture angle)
•
Integrated Standard Lens only
•
Add-On Lens (Wide Angle)
•
Add-On Lens (2 x Telephoto)
•
Add-On Lens (4 x Telephoto)
•
Add-On Lens (Macro)
TV43 (horiz. x vert.)
TV46 (horiz. x vert.)
34° x 25.5°
34° x 25.5°
45° x 34°
45° x 34°
17° x 12.7°
17° x 12.7°
8.5° x 6.3°
8.5° x 6.3°
7.8 x 4.1 mm
(0.31 x 0.16 in)
11 x 8.2 mm
(0.43 x 0.32 in)
Focus range IR camera
•
Integrated Standard Lens only
•
Add-On Lens (0.75X Wide Angle)
•
Add-On Lens ( 2X Tele)
•
Add-On Lens ( 4X Tele)
•
Add-On Lens (1.43X Macro)
152 mm (5.9 in) to ∞ (motorized remote focus)
152 mm (5.9 in) to ∞ (motorized remote focus)
406 mm (16 in) to ∞ (motorized remote focus)
2540 mm (100 in) to ∞ (motorized remote focus)
11 mm (0.42 in) (fixed fokus)
Focus range Visible Light camera
600 mm (23.6 in) - ∞ (fixed focus)
3.4. Dimensions
Figure 4: Front, Rear and Side View Dimensions of Standard ThermoView
®
Imager