There are three measurement mode positions associated with this switch:
Automatic Semi
Automatic Manual
In
Automatic
mode the Ti30 imager automatically adjusts the image to show the
lowest temperature value present in the thermal scene (MIN) and the highest temperature
value in the scene (MAX).
In
Semi-Automatic
mode the Ti30 imager keeps calculating the MIN limit (lowest
temperature value on the thermal scene) automatically.
In
Manual
mode the user can adjust
Gain
and
Level
manually.
See Selecting the Measurement Mode on page 58 for more information. The default position
is automatic mode.
Gain and Level Thumbwheels
The
Level Thumbwheel
control adjusts the median point of a given temperature
scale.
The
Gain Thumbwheel
control adjusts the temperature band or range around the
median
Level
point.
See Gain and Level Thumbwheels on page 61 for more information about these controls.