Figure 61. Editing Data Before Upload
Note
1:
Checking the Lock Emissivity or Lock Reflected Temperature Compensation will
prevent the operator of the imager from changing these values in the field. Pressing
the mode button on the imager from the Measurement mode will go immediately to
stored images in the imager (skipping emissivity and RTC modes completely).
Note
2:
If the need arises to change the order of images in a session (e.g. move the image
and information associated with it in location 31 to location 19), simply select the
image in location 31, click and drag the cursor to location 19 and release. The
image will be moved from location 31 to location 19 and the images between 19
and 30 will all be moved back one location. (Note: if you move an image to a
higher location number, the images between the new and old locations are all
moved forward one location.)
Note
3:
Alarm limits can be set during the image upload procedure for each image. In the
event that exceeding a particular temperature value is of interest (it may indicate
that a piece of equipment is nearing catastrophic failure for example), an alarm limit
can be set (either a high alarm, low alarm or both). If the temperature of the center
Jan 31st, 2005
- 71 -