spot exceeds (is higher than the high alarm limit or lower than the low alarm limit)
the limits set, the temperature display on the instrument will appear in BOLD RED
letters.
To upload the data to the imager, press
Upload
. The imager will begin to load data. Once
the data is loaded, you are ready to perform your inspections. Upon your return, you simply
place the imager in the docking station, and bring up the software. The new data is sent to
the proper destination and is ready for comparison with any earlier data.
Note
1:
It isn’t necessary to upload data to the imager in order to perform inspections.
However, it is recommended in order to keep consistency between inspections by
using the same parameters for each location to be inspected.
Note
2:
If you wish to delete all images currently stored in the imager, press the
Clear
images in imager
button.
By doing this all 100 locations on the imager will be restored with the factory
parameters. These default parameters are:
Location
name:
blank
Emissivity:
1.00
RTC:
OFF
Low alarm: 0°C / 32° F
High alarm: 250°C / 482° F
Comments:
blank