Maintenance, Calibration and Troubleshooting
Maintenance
4
4-1
Section 4
Maintenance, Calibration and Troubleshooting
4.1 Maintenance
The 943-227-15 Digital Preamplifier is designed to operate for extended periods of time with no
scheduled maintenance required. Operation may be verified by periodically actuating the check source
(provided with the detector and sampling assembly) and observing the response of the unit. If the
response varies by more than 50% of the normal value, further troubleshooting may be required. If a
problem develops, troubleshoot the unit per Section 4.3 and the drawings in Appendix B.
Whenever the preamplifier enclosure is opened, the condition of the internal desiccant package should be
checked. The
desiccant bag is provided in the enclosure to absorb moisture. The desiccant bags should
be checked for moisture absorption prior to placing into service, or whenever the electronics enclosure
cover is opened. Each bag contains a color indicator stripe that turns red when it should be replaced.
4.2 Electronic Calibration
Electronic Adjustments
There are no user adjustments available on the 943-227-15 Digital Preamplifier. The Programmable Gain
Amplifier and Auto Zero are factory adjustments, requiring the use of a factory test PROM and test fixture.
We strongly recommend the unit be returned to the factory for repair or adjustment.
Table 4-2 lists the electronic adjustments for the 943-227-15 digital preamplifier. The procedures for
making these adjustments are contained in the factory Calibration Procedure listed below. Note that
special test firmware is required to perform the electrical adjustments.
Table 4-1. Electronic Adjustments
Potentiometer
Adjustment
(943-227-15)
Tolerance
R11
Programmable Gain Amplifier Offset Adjustment
± 2 mV
R12
Auto Zero Set-Up Adjustment
± 10 mV
NOTE
Refer to the drawings in Appendix B to locate adjustment
and/or test points, if required.
The follow factory calibration and test procedures are listed for reference only.
Document
Description
TP943-227-15VL Test
Procedure,
Victoreen
Loop Communications
TP943-227-15
Test Procedure, RS232 communications
TP977-200-15M
Electrometer PCB Test Procedure
Summary of Contents for Victoreen 943-227-15
Page 2: ...Fluke Biomedical 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 www flukebiomedical com ...
Page 4: ... Blank page ...
Page 12: ... Blank Page ...
Page 21: ...943 227 15 Digital Preamplifier Operator Manual 3 3 ...
Page 25: ... Blank page ...
Page 31: ... Blank page ...
Page 33: ... Blank page ...
Page 37: ...943 227 15 Digital Preamplifier Operator Manual D 4 ...
Page 38: ...Appendix Cable Termination Instructions D D 5 ...
Page 39: ...943 227 15 Digital Preamplifier Operator Manual D 6 ...
Page 40: ...Appendix Cable Termination Instructions D D 7 ...
Page 41: ...943 227 15 Digital Preamplifier Operator Manual D 8 ...
Page 43: ... Blank page ...
Page 44: ... Blank page ...
Page 45: ...Fluke Biomedical 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 www flukebiomedical com ...