943-227-15 Digital Preamplifier
Operator Manual
1-4
Table 1-3. Preamplifier Serial Interface Board 977-210-10M Switch Settings
Switch Normal
Op.
Switch Function
SW1
Open
Microprocessor Reset switch, momentary contact
SW2
N/A
Not supplied
SW3-1
ON
OFF Selects 300 baud •
1/ON Selects 4800 baud (normal setting
)
SW3-2
OFF
OFF • (No function is implemented in the 977-210-10M)
SW3-3
OFF
OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for
factory PGA and Auto Zero set-up only)
SW3-4
OFF
OFF • (No function is implemented in the 977-210-10M)
NOTE
SW3 is a 4-position DIP switch on the 977-210-10M serial
interface board.
Table 1-4. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M
Jumper
Normal Setting Function
Alternate Setting Function
JMP1
A to B
Enables
Victoreen
Loop
(normal)
B to C
Enables RS232C
JMP2
N/A
Deleted, Not Used
N/A
Deleted, Not Used
Table 1-5. Jumper Positions for Preamplifier Electrometer Board 977-200-15M
Jumper
Normal Setting Function
Cal. Position Function
JMP1
A to B
Normal Operation
B to C
PGA Offset Adjust *
* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment
Summary of Contents for Victoreen 943-227-15
Page 2: ...Fluke Biomedical 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 www flukebiomedical com ...
Page 4: ... Blank page ...
Page 12: ... Blank Page ...
Page 21: ...943 227 15 Digital Preamplifier Operator Manual 3 3 ...
Page 25: ... Blank page ...
Page 31: ... Blank page ...
Page 33: ... Blank page ...
Page 37: ...943 227 15 Digital Preamplifier Operator Manual D 4 ...
Page 38: ...Appendix Cable Termination Instructions D D 5 ...
Page 39: ...943 227 15 Digital Preamplifier Operator Manual D 6 ...
Page 40: ...Appendix Cable Termination Instructions D D 7 ...
Page 41: ...943 227 15 Digital Preamplifier Operator Manual D 8 ...
Page 43: ... Blank page ...
Page 44: ... Blank page ...
Page 45: ...Fluke Biomedical 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 www flukebiomedical com ...