Guidance and Manufacturer’s Declaration – Immunity
The FLYP pel200m is intended for use in the electromagnetic environment
specified below. The customer or user of the FLYP pel200m should ensure that it
is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment – Guidance
ESD:
IEC61000-4-2
±8kV Contact
±2KV, ±4kV,
±8kV, ±15kV Air
±8kV Contact
±2KV, ±4kV,
±8kV, ±15kV Air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be 30%.
EFT:
IEC61000-4-4
+/- 0.5 KV L1
and L2
+/- 1 KV L1
and L2
+/- 0.5 KV L1
and L2
+/- 1 KV L1
and L2
Mains power quality should be
that of a typical commercial or
hospital environment.
Surges:
IEC61000-4-5
+/- 0.5 KV L1
and L2
+/- 1 KV L1 and L2
+/- 0.5 KV L1
and L2
+/- 1 KV L1 and L2
Mains power quality should be
that of a typical commercial or
hospital environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
Harmonics:
IEC61000-
3-2
Not
Applicable
The FLYP pe1200m is suitable for use in all
establishments, including domestic, and
those directly connected the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Flicker:
IEC61000-
3-3
Not
Applicable
Guidance and Manufacturer’s Declaration - Emissions
The FLYP™ pe1200m is intended for use in the electromagnetic
environment specified below. The customer or user of the FLYP pe1200m
should ensure that it is used in such an environment.
EMC Tables. The following tables are provided in accordance with IEC
60601-1-2: 2014, ed. 4.0.