Immunity Test
IEC 60601 Test Level
Compliance Level
Conducted RF:
IEC61000-4-4
Radiated RF:
IEC61000-4-3
3 Vrms 0.15-80 MHz
6 Vrms in ISM and
Amateur radio bands
10 V/m
80 MHz - 2.7 GHz
1 kHz @ 80% AM
3 Vrms 0.15-80 MHz
6 Vrms in ISM and
Amateur radio bands
10 V/m
80 MHz - 2.7 GHz
1 kHz @ 80% AM
Electromagnetic Environment – Guidance
Portable and mobile communications equipment should be separated from the
FLYP pe1200m by no less than the distances calculated/listed below:
Guidance and Manufacturer’s Declaration – Immunity
The FLYP™ pel200m is intended for use in the electromagnetic
environment specified below. The customer or user of the FLYP pel200m
should ensure that it is used in such an environment.
D=(3.5/V1)(Sqrt P)
150kHz to 80 MHz
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
where P is the max. power in watts and D is the
recommended separation distance in meters.
Field strengths from fixed transmitters, as determined by
an electromagnetic site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment
containing a transmitter.