36
Error Messages
MESSAGE
WHAT IT MEAN
WHAT TO DO
Appears when the
batteries cannot provide
enough power for a test.
Replace the batteries
immediately.
Appears when a used test
strip is inserted.
Repeat the test with a new test
strip.
Problem in operation.
Repeat the test with a new test
strip. If the meter still does not
work, please contact the customer
service for assistance.
Appears when test strip is
removed while counting
down, or insufficient
blood volume.
Repeat the test with a new test
strip. If the problem persists,
please contact the local customer
services for help.
Appears when the wrong
code strip is inserted or
other coding errors.
Make sure to insert the right code
strip for the meter.
Check if the code numbers on the
code strip and the strip vial label
are the same.
Expired code strip.
Repeat the test with a new code
strip.