background image

For questions on how to use the FORA medical devices, please contact the

FORA Customer Service Line

 at 

1-888-307-8188

Troubleshooting

3G Cellular Data Transmission

1. The RF indicator will flash a orange light when the 3G service is searching for the cellular connection.  
2. The RF indicator will turn a solid blue when the connection is found and it starts the data transmission.

Troubleshooting the FORA D40g

Error Signal Codes
If there is an error, the device will show ERR in the screen along with a code. Refer to your manual for the error code listing. If you are 
having trouble establishing a signal, please move the device into a more open area, or closer to a window.

2

1

RN Staffing Solutions, LLC. New York        |        Toll Free: 718-720-7242        |        www.rnstaffingsolutions.com/

For questions on the 24/7 HealthView, please contact RN Staffing Solutions via phone at 

718-720-7242

 or via email at 

rnstaffingsolutions@gmail.com.

 
For questions on how to use the FORA medical devices, please contact FORA Customer Service Line at 

1-888-307-8188

Error Code

Err 1
Err 2

Appears when the batteries cannot provide 
enough power for a test.

Replace or recharge the 
batteries immediately. 

PCL Err 30

Unable to connect to the service.

PCL Err 40
PCL Err 41
PCL Err 42
PCL Err 43
PCL Err 44
PCL Err 45
PCL Err 48
PCL Err 49

No response from the server.

What It Means

Solution

Err 3

Err 4
Err 5

No pulse rate detected or the pulse rate is too 
low.

Cuff pressure is too high.

Inflation or pressure error.

Service error.

Failure connecting to the service.

Connection overtime.

Command failure.

Data upload error.

Problem connecting to the service.

Turn the device off and on again 
to reestablish a signal. For a 
better signal, move the device 
into a more open area or closer to 
a window. If the problem still 
persists, please contact customer 
service for assistance.   

Refit cuff tightly and correctly. 
Relax and repeat the measure-
ment.  If the error still remains, 
please contact customer service 
for help.

Summary of Contents for D40g

Page 1: ...D40g Blood Pressure and Blood Glucose Monitoring System Quick Start Guide EN ...

Page 2: ...0g Meter Information Intended for single patient use only Meter will switch off automatically if left idle for 3 minutes Meter should transmit data automatically For more information on manual data transmission or signal quality check please refer to the user manual Blood Glucose Strips Strips have an expiration date Please check strip container for the expiration date before using Strips expire 9...

Page 3: ...wer button All the LCD symbols will appear Then the cuff will begin to inflate automatically 7 The heart symbol will flash when a pulse is detected during the inflation 8 After the measurement the monitor displays the systolic pressure diastolic pressure and pulse rate Performing a Test Blood Glucose 1 Wash your hands 2 Insert lancet into FORA lancing device and close 3 Insert test strip into the ...

Page 4: ...ter handling the meter lancing device and test strips to avoid contamination This device has been validated to withstand up to 6 000 cleaning and disinfecting cycles using the recommended disinfecting wipes See the user manual for more information For additional information on the D40g please refer to the user manual FAQ How does the D40g work The D40g measures the systolic and diastolic blood pre...

Page 5: ...ine at 1 888 307 8188 Error Code Err 1 Err 2 Appears when the batteries cannot provide enough power for a test Replace or recharge the batteries immediately PCL Err 30 Unable to connect to the service PCL Err 40 PCL Err 41 PCL Err 42 PCL Err 43 PCL Err 44 PCL Err 45 PCL Err 48 PCL Err 49 No response from the server What It Means Solution Err 3 Err 4 Err 5 No pulse rate detected or the pulse rate i...