1.5 Location
Locate the freezer on a firm, level surface in an area with an ambient temperature
between 18°C and 32°C. Provide ample room to reach the mains disconnect switch
(power switch) located on the rear of the freezer.
For proper ventilation and airflow, a minimum of 5" of clear space
is required around the freezer.
1.6 Connecting the Alarm Battery
NOTE:
The battery must be connected upon start-up so that the alarm will activate
during an Overtemperature condition.
1.
To gain access to the battery, remove the grill on the front of the freezer. The grill is
attached to the freezer by friction plugs on each of the four corners and is readily
pulled off. The battery is the rectangle fastened by Velcro to the compressor
compartment frame and has a connector with red and black wires. Another red and
black wire set with connector is secured to the wiring harness passing through the
area of the freezer. Join the two wire connectors and replace the grill.
2.
When the battery is connected, the
OVERTEMP
light will flash and the alarm will
sound. Turn the N
ORMAL/STANDBY
switch to the
STANDBY
position. This will
silence the alarm. The alarm will remain silent until the unit is below alarm
setpoint. The alarm setpoint is factory set for -65°C. If another setpoint is required,
see Section 2.2.