Installing the FP S-Compact
27
The following types can be used for the mains cable:
– Flexible: e.g. H05VV-F 3G 0.75 mm²
– Rigid: e.g. H07 VV-U 3 G 1.5 mm².
Connect the phase (L) to the screw terminal with the ‘L’ mark-
ing. Please use wire end ferrules (pin length 6 mm / 0.24").
Connect the neutral line (N) to the screw terminal with the ‘N’
marking. Please use wire end ferrules (pin length 6 mm / 0.24").
Attach a ring eyelet (M4) to the end of the protective earth
wire.
Screw the protective earth wire onto the connection point
marked with the earthing symbol. Use the copper screw (M4)
provided to fasten it on as well as a toothed washer (M4
DIN6797A) above and below the ring eyelet.
Tighten the clamping screws to a torque of 0.5 Nm (0.37 ft lb).
Fitting the protective cover
Pick up the protective cover with both grips and set the
protective cover down as shown in the illustration.
Press the protective cover onto the holder.
The snap hooks latch into place. The protective cover is then
correctly fitted again.