Part No.16894 Rev. G March 2014
- 10 -
3000 Series Cabinets
TROUBLE SHOOTING
FAULT
POSSIBLE CAUSE
REMEDY
Doors are not sliding smoothly
Door not in track
Install door correctly in track
Debris in track
Clean door tracks
(see cleaning)
Lack of lubricant
Apply food grade lubricant to
door track
Cabinet does not operate/start
The mains isolating switch on
the wall, circuit breaker or
fuses are off at the power
board
Turn isolating switch circuit
breaker or fuses on
The thermostat is in the OFF
position
Turn control knob to desired
temperature
Internal fuse blown
Have wiring checked and
replace fuse
(5A Slow Blow)
Cabinet does not reach
temperature
One or more doors is open
Close doors and re-test
temperature after thirty
minutes
Thermostat setting disturbed
Re-set thermostat and re-test
after thirty minutes
Thermostat is faulty
Have thermostat replaced
An element is blown
Have the element replaced
Cabinet lights not working
The light switch is OFF
Turn light switch ON
Halogen lamp has failed
Replace lamp
Internal fuse blown
Have wiring checked and
replace fuse
(5A Slow Blow)
Aluminium parts corroded
Caustic detergent damage
Order replacement parts
Service
Personnel Only
The table entries in
italics
indicate actions to be taken only by qualified Service
Personnel.