- Pressure

The Multimeter can only be used with the specified
pressure sensor accessories (see Table above).

- Calibration of pressure sensor:

The pressure sensor must be connected to the in-
strument during the calibration procedure. Press the
Select button once, and the message: 

ProbCal

will

be shown in the display. Press Select again, and the
message: 

1.000 cF 

will be shown. The scale factor

for the pressure sensor can now be selected using
the Adjust buttons. The new selection is confirmed
by pressing the Select button, and the display
shows the message: 

0.0 Ofs

. It is now possible to

adjust the zero offset for the pressure sensor by
means of the Adjust buttons. Confirm the selection
by pressing Select. The instrument will then change
to the measurement mode and show the current
pressure reading. Note that the calibration factors
are stated on a label on each pressure sensor.

- Menu:

This option is intended for use with special facilities,
e.g. for connection to a PC. In the current version of
the Multimeter this option has not been implemen-
ted.

Technical data

Supply voltage:

/-10%

Power consumption: 20 W

Front panel display:

4-digit 7-segment LED, 
45 mm hih

Units display:

7x15 LED matrix

Bagdisplay:

4-digit 7-segment LED, 
13 mm high 

Fuse for current

mA-range: 2A (slow) 
(4090.21)

Current limiter:

A-range - electronic limit with
warning display

Fuse for grid:

315mA (slow) (4090.36)

(Located in fuse holder on back panel)

®

Measurement

Range

Accuracy

Input impedance

DC Voltage

0 – 500V

0.5% + 1 digit

10M

Ω

AC Voltage

20mV – 500V

10Hz – 1kHz: 1% + 1 digit

10M

Ω

1kHz – 2kHz: 2% + 2 digit

2kHz – 5kHz: 5% + 2 digit

5Hz – 10kHz: 10% + 2 digit

DC Current

0 – 10A

0.5% + 1 digit

10

Ω

; 0.1

Ω

; 0.01

Ω

AC Current

20µA – 10A

10Hz – 10kHz 1% + 2 digit

Resistance

0 – 10M

Ω

1% + 1 digit

Frequency *

0 – 100 kHz

0.5% + 1 digit

10M

Ω

Temperature

-200°C – 1370°C

0.1% + 1 digit

Pressure

0 – 7000 hPa

0.1% + 1 digit

pH

0-14

0.1% + 1 digit

>10G

Ω

* Frequency measurements up to 10 kHz require at least 1 volt; from 10 kHz to 100 kHz require at least 10 volts.