995066 - Revision A
Page 6 of 41
4.0 – Test Procedure
Below is the full testing procedure required to complete an Annual Stand Aid Service.
The SWL (180Kg) is an essential load for the testing of the Stand Aid and a mode of transportation to
transport the weights to the hoist is required.
1.
Position the Stand Aid on a flat and level surface and away from any obstacles.
2.
Using the handset raise the Upper Handle Assembly to the upper limit. Listen for unusual
noise.
3.
Lower the Upper Handle Assembly to the bottom limit and ensure that the actuator lowers
smoothly without resistance.
4.
Using the handset, open the legs to their widest position.
5.
Apply the brake to the rear castors.
6.
Using a calibrated 180kg weights. Position it perpendicularly under the Upper Assembly.
7.
Attach the weight stack to the Upper Assembly using lift straps.
8.
Raise the Weight 2/3 the lifting range. Be vigilant for unusual noises while raising the weight.
9.
Deactivate the brakes and close the legs using the handset.
10.
Drive the Stand Aid forward and backwards one metre. Make sure that the castors rotate
and roll correctly.
11.
Activate the rear castors brakes.
12.
Using the handset begin to lower the weight. Before reaching the floor, press the E-STOP
button on the control box to check the safety device is working correctly.
13.
Begin to twist the Red manual E-lower device positioned at the top of the actuator until
weight has reach the floor.
14.
Detach the weight from the Upper Handle Assembly.
15.
Twist the E-STOP Button in a clockwise motion until it returns to its natural position.
16.
Once weight is removed continue lowing the Upper Handle Assembly down to the bottom
limit.
17.
Apply 45kg of weight directly to the Foot Tray to test the Foot Tray is still rigid and does not
bend significantly under load.
18.