The battery is fully charged
The battery is 80% charged
The battery is 60% charged
The battery is 40% charged
The battery is 20% charged
The battery is nearly empty
Battery life is at least 30 minutes remaining
►A pre-alarm is triggered (see
on page 80)
/
(in yellow)
The battery is empty (approx. 10 minutes left), an alarm is triggered (see
on page 80)
(in red)
INFORMATION
■
If battery is failing, do not use the device. Return device to Fresenius Kabi
sales representative as soon as possible.
■
Battery replacement must be performed by qualified and trained technical
personnel in compliance with the technical manual and procedures.
■
Under normal condition of use, battery life might be reduced from 24 hours
to 20 hours by the end of the third year of use.
5.1.3 Battery precautions
The device uses a Lithium-ion rechargeable battery.
The following actions may cause leakage, overheating, smoke, explosion or fire, which
could result in deterioration of performance, failure, damage to the equipment or injury to
the user:
■
Incorrect handling of a Lithium-ion battery.
■
Replacement of the battery by inadequately trained personnel.
INFORMATION
■
Do not replace with a battery other than the one provided by
Fresenius
Kabi
.
■
Do not use the pump without the battery connected.
■
Do not disconnect the battery when the device is operating on mains power
supply or battery power. Disconnect the power cable and power off the
device before disconnecting the battery.
■
Do not incinerate or place near a flame.
■
Do not drop, crush, puncture, modify or disassemble the battery.