Do not use a battery that is severely scratched or damaged.
■
Do not short the terminals.
■
Do not expose to high temperatures or very low temperatures:
refer to the operating conditions for use, and the storage and transport
conditions.
■
Do not try to charge or discharge the battery outside the device.
■
For more information on replacing the battery, refer to the technical manual.
5.2 Switch on
INFORMATION
■
Before using the pump, proceed to
on page 77.
■
When using a pump on patient requiring special attention, ensure that a
backup pump is available for immediate use especial in homecare
environment.
■
Before switching on the pump, install holder and pump, see
on
page 19.
■
The pump can operate using the battery, however, we recommend that the
pump be connected to a power supply as often as possible during use to
ensure that the battery remains charged.
■
When the pump is connected to the power supply, check that the power
supply indicator lights up green
, and that the power cable and the
wall plug are accessible.
To switch on the pump, press the ON/OFF key
until the screen light up.
At start-up, the device displays the product name and logo, the version number of the pump's
software. An auto-test checks the functionality of the module.
During the 2-second autotest:
■
red, yellow and green LEDs blink;
26