4.
Press
to start the infusion.
Accessing to information screen is described in
on page 33.
6.4.5 Intermittent mode
INFORMATION
■
The target volume of one cycle can be set from 1 mL to 9999 mL with a
default value of 50 mL.
■
The number of infusion cycle can be set from 2 to 99 with a default value
of 3.
■
The target duration of one cycle can be set from 15 min. to 24 h with a
default value of 8 h.
Intermittent mode enables cyclic infusion at an adjustable flow rate, volume and time.
1.
Select the intermittent mode and press
(
) to enter. For mode selection,
see
on page 44.
50