2.
For level of occlusion pressure setting, press
to select the low level
or press
to
select the high level
of occlusion pressure (high by default).
NOTE:
Low level = an occlusion alarm will be triggered for a pressure value of 487.5 mmHg ± 187.5
mmHg in the administration set; High level = an occlusion alarm will be triggered for a pressure value
of 787.5 mmHg ± 262.5 mmHg in the administration set.
3.
Press
to select the single bubble setting. Press
or
to adjust the value.
NOTE:
Air in line alarm will be triggered when the detected single bubble volume is ≥ to the
customized volume value. The single bubble setting can be adjusted from 0.125 mL to 0.75 mL, with
step of 0.125 mL (default setting: 0.125 mL).
62