4.
Press
to go down to select the cumulated bubbles volume setting. Press
or
to switch
between volume and duration settings and press
to adjust the value of the selected setting.
NOTE:
Air in line alarm will be triggered when the detected cumulated bubbles volume is ≥ to the
customized volume value throughout the customized duration. The adjustment range of cumulated
volume is from 0.25 mL to 0.75 mL by step of 0.125 mL, the adjustment range of time is 1 min, 5 min
, 10 min, 15 min, 20 min, 25 min, 30 min (cumulated bubbles volume setting is 0.5 mL / 30 min by
default).
5.
Validate the settings by press
(
) and pump returns to the infusion programming screen.
INFORMATION
■
The value of single bubble means the minimum size of the single bubble
that can be detected by the pump, the detection will last until the end of
infusion.
■
The cumulated bubbles volume detection is used to detect the total volume
of air bubbles during the set duration, the detection will last until the end of
infusion.
6.6.2 End of infusion pre-alarm
This function enables to customize alarm to notify approaching end of infusion.
Ambix nova enables pre-alarm before the end of infusion for Continuous mode, Intermittent
mode and Ramp mode.
1.
Under advanced settings menu (see
on page 60 to access),
select the infusion pre-alarm menu
and press
(
) to enter. Press
to deactivate or press
to activate (active by default).
63