3.
Press
to select the KVO flow rate value and adjust it by pressing
(the
parameter can only be selected when KVO is activated).
NOTE:
KVO rate setting range is from 1 mL/h to 10 mL/h, the default value is 2 mL/h.
4.
Press
(
) to validate the setting. Pump returns to the infusion programming
screen.
INFORMATION
■
Check KVO settings before start infusion, make sure KVO rate is appropriate
for the patient.
■
KVO status is displayed on the infusion programming screen and infusion
information screen (except in flow rate only mode) but cannot be adjusted.
Under flow rate only mode, KVO is only functional when delayed start
function is activated.
■
After activation, KVO is functional during all infusion interruptions:
–
Delayed start
–
Pause periods of an intermittent infusion cycle
–
End of infusion
67